Braun Stimuplex HNS 11 HNS 11 User Manual Page 5

  • Download
  • Add to my manuals
  • Print
  • Page
    / 28
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 4
In the operating mode, the LCD display shows the preselected
stimulation current set on the controller between 0.00 and
approx. 1.00 mA (in the 1 mA range) or from 0.00 to approx.
5.00 (in the 5 mA range). In the operating mode, the Stimu-
plex
®
HNS 11 shows the stimulation current actually flowing
via the stimulation electrodes through the tissue. A + sign on
the LCD display signals this setting. The operating mode is
switched by pressing the mA button.
3. Select operating mode . Set the controller to 5 mA max. and
try to get a connection with the insulated, black 2 mm male plug
with a part of the red clamp. Compare – by briefly switching to
operating mode – the measured value with the set value (ac-
tual value and set value). These two values must be the same –
with shorted cables. Reset to 1 mA max. and compare these two
values (see Measuring Accuracy in Appendix A, if neccessary).
A
~
symbol appears next to the symbol in the LCD display
(Display) and the pulsed LED display (orange light-emitting di-
ode) blinks according to the stimulation frequence.
A "+" sign appears
in front of each measurement displayed.
When a current is set at
O, the
orange LED starts to blink at 0.02
mA, when 1 mA max. is set, and at 0.10 mA when 5 mA max. is
set.
4. When the stimulation cable is disconnected (when is set),
+0.00 appears in the display and the
~
symbol disappears. The
pulsed LED display is no longer lit.
If the "BAT" symbol appears in the display, then the battery
voltage is getting too low. Current application can be complet-
ed without a problem, but exchanging the battery afterwards
should be considered. Press the "BAT" button to display the re-
maining battery voltage (in volts).
When the battery voltage is approx. 6 volts, three horizontal
dashes preceded by the "+" sign (or four dashes) appear in the
display in addition to the "BAT" symbol. In this case, the stimu-
lator shuts off automatically.
To avoid having to abort a stimulator treatment, you should
replace the 9 volt block battery as soon as the battery volt-
age drops to 6.5 volts.
5. To switch off the nerve stimulator, turn the control knob anti-
clockwise to position O.
To conserve the battery always switch off the Stimuplex
®
HNS 11
after use. If you do not use it for a long period of time, remove
the battery to prevent battery leakage.