8
9
IMPORTANT
Read the operating instructions.
Classication:
• Internally powered equipment
Type BF equipment
• IPX0
Not suitable for use in the presence of ammable anæsthetic mixture with air,
oxygen or nitrous oxide
• Continuous operation with short-time loading
This product conforms to the provisions of the EC directive 93/42/EEC (Medical
Device Directive). This device conforms to the following standards:
EN 60601-1: 1990 + A1: 1993 + A2: 1995 + A13: 1996 General requirements for
safety.
EN 60601-1-2: 2001 - Electromagnetic compatibility requirements and tests.
EN1060-1: 1995 + A1: 2002 - Non-invasive sphygmomanometers- General
requirements .
EN1060-3: 1997 + A1: 2005 - Non-invasive sphygmomanometers- supplementary
requirements for electromechanical blood pressure measuring systems.
EN1060-4: 2004 - Non-invasive sphygmomanometers- Test procedures to
determine the overall system accuracy of automated non-invasive
sphygmomanometers.
Portable and mobile RF communications equipment can aect MEDICAL
ELECTRICAL EQUIPMENT.
Please do not dispose of the product in the household waste at the end of its
useful life. Disposal can take place at your local retailer or at appropriate collection
points provided in your country.
Guarantee
Kaz grants 2 years guarantee on the product commencing on the date of purchase. Within the
guarantee period we will eliminate, free of charge, any defects in the appliance resulting from
faults in materials or workmanship, either by repairing or replacing the complete appliance as we
may choose.
This guarantee extends to every country where this appliance is supplied by Kaz or its
appointed distributor.
This guarantee does not cover: damage due to improper use, normal wear or use as well as defects
that have a negligible eect on the value or operation of the appliance. The guarantee becomes
void if repairs are undertaken by unauthorised persons and if original Braun parts are not used.
To obtain service within the guarantee period, hand in or send the complete appliance with your
sales receipt to an authorised Kaz Customer Service Centre (See package insert for a local Service
Center)
UK only:
This guarantee in no way aects your rights under statutory law.
What to do if …..
Problem Reason Solution
Heart Rate
Symbol
Appears in the measurement
condition and ashes when
pulse is detected.
Measurement in progress, remain quiet.
Low Battery
Indicator
Appears when the battery
voltage is excessively low or
the positions of the
batteries are incorrect.
Replace all four batteries with new ones. Insert
the batteries in the correct positions.
Be aware of +/- positions.
Measuring
Error
Appears when the
accurate blood pressure and
pulse could not be
obtained.
Press “Start/Stop” button again and remeasure.
Check if cu is wrapped according to
instructions.
Check that there are no kinks in tube.
Check palm if exerting eort.
Check if talking or moving during
measurement.
Check if posture is correct.
Specications
Method of measurement Oscillometric
Model number BP4600
Range of
measurement
Pressure 0~300 mmHg
Pulse 40~199 beats/minute
Accuracy Pressure +/- 3 mmHg
Pulse +/- 5% Max.
Ination Deluxe Automatic
Exhaust Automatic exhaust valve
Display Liquid Crystal Display – Systolic, Diastolic, Pulse rate
Sets of memory 32 sets
Cu size • Small/Medium Cu = 23~33 cm arm circumference
• Large/XLarge Cu = 33~43 cm arm circumference
Operating temperature +10° ~ +40°, less than 85% R.H.
Storage temperature -20° ~ +70°, less than 85% R.H.
Unit weight 290g (without batteries)
Power supply Alkaline Battery: 4 x AA (LR6)
Battery life 300 times measurement (twice a day measurement)
Auto power o Whenever not used for 1 minute
Accessories 4 batteries, arm cu with tube, instruction manual, pouch, carton box