Sterile container system
Aesculap Sterile Technology
5.3 Manual cleaning/disinfecting
¾ Inspect visible surfaces for residual contamination after manual
cleaning/disinfecting.
¾ Repeat the cleaning process if necessary.
Manual cleaning and wipe disinfection
D–W: Drinking water
FD–W: Fully desalinated water (demineralized)
RT: Room Temperature
Stage I
¾ Wipe off gross debris prior to cleaning.
Stage II
¾ Dry the product with lint-free tissue or medical-quality filtered
compressed air.
Stage III
¾ Carry out wipe disinfection.
Stage IV
¾ When using aldehyde-free disinfectants:
After the specified exposure time (at least 1 minute), rinse the
disinfected surfaces under running FD water.
¾ Allow water to drip off for a sufficient length of time.
Stage V
¾ Dry the product with lint-free tissue or in a heating cabinet.
Stage Step T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemical
I Cleaning RT
(cold)
--DW-
II Drying RT - - - -
III Wipe disinfection -
>1- -I
Alcohol Denat. 70 %
(ethanol B|BRAUN)
II
Aldehyde-free surface
disinfectant (Stabimed)
IV Final rinse RT
(cold)
- - FD–W not required Rinse off cleaning
chemicals; do not leave
residues
V Drying RT - - - - -