Braun 42194 User Manual Page 11

  • Download
  • Add to my manuals
  • Print
  • Page
    / 84
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 10
9
5.4 Mechanical neutral or mild-alkaline cleaning and thermal disinfection
Machine type: Single-chamber washer/disinfector without ultrasound
D–W: Drinking water
FD–W: Fully desalinated (demineralized) water
¾ Check visible surfaces for residues after mechanical cleaning/
disinfecting.
¾ Repeat the cleaning process if necessary.
Note
Ultrasonic treatment can be applied in manual processing as well as in
mechanical processing. Temperatures of up to 120 °C are permissible for
mechanical drying with hot air.
5.5 Inspection, maintenance and checks
Note
The sterile containers may be tested and repaired only by persons with the
appropriate training, expertise or experience.
The seals in the container lids were tested according to DIN EN 868-8,
Annex G, for a minimum service life of 5 000 sterilization cycles.
¾ In case of visible damage, change the seal immediately.
¾ Have the lid sent to Aesculap, see Technical Service.
¾ If necessary, lightly lubricate moving metal parts, e.g. latch hinges,
with sterilizable and steam-permeable maintenance oil (e.g. STERILIT®
oil spray JG600 or STERILIT® I drip feed oiler JG598).
Stage Step T
[°C/°F]
t
[min]
Water
quality
Chemical
I Prerinse <25/77 3 D–W -
II Cleaning 55/131 10 FD–W Neutral:
BBRAUN HELIMATIC CLEANER
neutral, pH-neutral,
process solution 0.5 %
Mild-alkaline:
- Concentrate:
pH = 9.5
< 5 % anionic tensides
- 0.5 % solution
Alkaline:
Processing possible at pH up to 10.5, provided the cleaning
agent is manufacturer-approved for container cleaning.
Note
Application of mild-alkaline or alkaline cleaning agents
may cause discoloration of colored anodized aluminum,
which however will not affect the functionality of the
product.
III Intermediate rinse >10/50 1 FD–W Especially for PrimeLine it must be ensured that the
surface is rinsed without leaving any residues.
IV Thermal disinfecting 90/194 5 FD–W Other process parameters may be feasible with agreement
by the hospital hygienist.
V Drying - - - According to mechanical cleaning program
Page view 10
1 ... 10 11 12 ... 84

Comments to this Manuals

No comments